Work

Detecting excessive and insufficient breathing effort with respiratory surface EMG

Year: 2025
Type: article
Abstract: During mechanical ventilation risks for excessive and insufficient breathing effort directly determine patient outcomes through distinct injury mechanisms. Respiratory surface electromyography (sEMG) holds potential for monitoring breathing effort, however measurement of electrical activity in comparison to conventional pressure measurement exhibits systematic differences in timing and amplitude of inspiratory activity. In this article influences of neuromechanical delay (NMD) and coupling (NMC) on detection performance using sEMG compared to pressure measurements are investigated. Results indicate excellent performance if sEMG measurements are corrected for both NMD and NMC. However, using generalized values for NMD and NMC do not achieve improvements, highlighting the importance of individualized calibration.
Language: English
Field: Medicine
Open Access status: green
OpenAlex ID: W7135384319
FWCI:
Cites:
Cited by:
Related to: