Year: 2016
Type: article
Abstract: Ceramic thermoluminescence detectors (TLDs) are low-cost devices with high spatial resolution and very easy to use.The basic characteristics of alumina-based ceramic TLDs (Chiba Ceramic Mfg.Co., Ltd.A8) in response to X-ray and charged particle beams (H, He, C, Ne, Ar) have been reported in an effort to apply TLDs as a QA/QC tool for charged particle therapy.The glow curve has two peaks, and the main peak is located at 148 °C and 694 nm.For each irradiation, four dose levels were given: 0.5, 1.0, 2.0, and 5.0 Gy in water.The main peak showed good linearity and repeatability for X-rays.However, thermoluminescence efficiency decreased slightly with increasing irradiation dose with charged particle irradiation.The thermoluminescence efficiency of the main peak showed a very small dependence on linear energy transfer (LET) compared with common TLDs.This ceramic TLD may be applicable as a verification tool in a limited LET range such as proton therapy.
Source: Sensors and Materials
Language: English
Subfield: Radiation
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: gold
Estimated APC paid: $748
OpenAlex ID: W2530741806