Uncovering the origin of impurities in supercritical fluid chromatography coupled to high-resolution mass spectrometry

Year: 2026
Type: article
Abstract: Maintaining low background levels is essential for analytical instruments, such as in supercritical fluid chromatography (SFC) coupled to high-resolution mass spectrometry (HRMS). An approximate tenfold increase of the background was observed in our laboratory when a new CO2 cylinder with a dip tube was installed on a SFC-HRMS system. Most analyte peaks were obscured. This study aimed to find the cause of the elevated background. Accordingly, the entire system was cleaned and each module, including the ion source, back pressure regulator, column, autosampler, solvent pumps, CO2 pump, and cylinder, were assessed. The troubleshooting suggested that the source of the contamination was the CO2 cylinder. A statistically significant (p > 0.05) elevation in the background was observed when liquid CO2 was used as mobile phase. The ions present above normal background levels were found to have m/z of 492.34388, 495.29340, 371.24343, 490.33806, and 490.33810. Annotation of the most abundant ion (m/z of 490.3310) based on m/z, isotope pattern, fragmentation pattern, and collision cross section, suggested structures resembling derivatives of sphingo- and glycolipids. A possible source could be degradation products formed during bioethanol production, the process from which the CO2 in the cylinder originates. Noteworthily, another SFC-HRMS system that uses condensed gaseous CO2 instead of liquid CO2, has so far not resulted in any contamination-related issues. To prevent contamination in sensitive SFC-HRMS systems, it is essential to assess the purity of the CO2 before use, especially if liquid CO2 from a cylinder with dip tube is used.
Institution: Lund University
Language: English
Subfield: Spectroscopy
Field: Chemistry
Open Access status: hybrid
Estimated APC paid: $4,040
OpenAlex ID: W7154395281
FWCI:
Cites: 15
Cited by:
Related to: