Magnetic Resonance Imaging Safety in Patients With Conditional Mixed‐Manufacturer Devices
Year: 2026
Type: article
Study design: Observational Study
Abstract: BACKGROUND: Magnetic resonance imaging (MRI) is a critical diagnostic tool known for its non-invasive nature and high-resolution imaging capabilities. However, its safety in patients with cardiac implantable electronic devices (CIEDs), particularly those with mixed-lead systems, remains a significant clinical concern. This report investigates the safety and efficacy of MRI in such patients, where generators and leads originate from different manufacturers and are not officially deemed MRI-conditional. METHODS: A retrospective, non-randomized single center study was conducted involving 16 patients with mixed company lead CIEDs that obtained 23 MRIs. All patients underwent clinically indicated 1.5 Tesla MRI scans between July 2021 and August 2025. RESULTS: Pre- and post-scan measurements included lead impedance, sensing thresholds, and battery voltage, along with monitoring for clinical complications such as arrhythmias, syncope, or cardiac arrest. No adverse effects or device malfunctions were observed, and all key parameters remained stable post-scan. CONCLUSION: This study shows the safe use of MRIs for patients with mixed-lead CIEDs.
DOI: 10.1111/pace.70297
Institutions: Abbott (United States), University of Southern California
Language: English
Subfield: Cardiology and Cardiovascular Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: hybrid
Estimated APC paid: $4,170
OpenAlex ID: W7162117103