Year: 2024
Type: article
Abstract: The pattern electroretinogram (PERG) is a localized retinal response evoked by a contrast-reversing pattern, usually a black and white checkerboard, which provides information about macular and retinal ganglion cell function. This document, from the International Society for Clinical Electrophysiology of Vision (ISCEV; www.iscev.org ) presents an updated and revised Standard for clinical PERG testing. This replaces the 2013 and all earlier versions. Minimum protocols for basic PERG stimuli, recording methods and reporting are specified, to promote consistency of methods for diagnosis and monitoring purposes, while responding to evolving clinical practices and technology. The main changes in the updated ISCEV Standard for clinical PERG include expanded guidance about large stimulus fields, stimulus parameters for simultaneous PERG and pattern visual evoked potential recording, baseline drift correction, and use of consistent ambient room lighting. These changes aim to provide a clinically relevant document about current practice which will facilitate good quality recordings and inter-laboratory comparisons.
Source: Documenta Ophthalmologica
Institutions: Great Ormond Street Hospital, Great Ormond Street Hospital for Children NHS Foundation Trust, University College London, University of Freiburg, University Medical Center Freiburg, University of Illinois Chicago, Ljubljana University Medical Centre, State University of New York, SUNY College of Optometry, Moorfields Eye Hospital NHS Foundation Trust, Moorfields Eye Hospital
Language: English
Subfield: Ophthalmology
Field: Medicine
Domain: Health Sciences
Open Access status: hybrid
Estimated APC paid: $3,760
OpenAlex ID: W4392853687