Practicality of Using Pressure Sensors and Accelerometers to Quantify Hand Orthosis Compliance at Home
Year: 2026
Type: article
Abstract: Orthosis compliance monitoring provides insights into effective orthosis design and user wear time. Frequently, patient reports of orthosis use are subjective and often result in overestimation of compliance. Therefore, a tool to objectively observe whether patients wear their orthoses as instructed is vital. This study assessed the real-world practicality of using an objective compliance monitoring device with a hand orthosis. A device consisting of a pressure sensor and accelerometer was tested by ten healthy volunteers who wore a hand orthosis daily and completed a diary of their wear time and activities for a week. Sensor data obtained from the compliance monitoring device were analysed to discern each user's orthosis wear time. Differences between estimated wear time and actual wear time were insignificant. Pressure-based wear time estimations had a specificity of 99.3 ± 0.7% and a sensitivity of 80.3 ± 19.2%, whilst acceleration-derived estimations had a specificity of 94.5 ± 6.4% and a sensitivity of 73.2 ± 15.8%. This study demonstrated that orthosis compliance can be monitored outside the laboratory, and, furthermore, this device offers insights into the intensity and frequency of a user's activities and has the future potential to monitor orthosis fit and forces applied to affected joints using pressure.