Abstract: INTRODUCTION: Generalized myasthenia gravis (gMG) is a chronic autoimmune disorder that causes muscle weakness. This human factors (HF) validation study aimed to demonstrate that use of a prefilled syringe with needle safety device (PFS-SD) or autoinjector (AI) is safe and effective for patients with gMG, caregivers, and healthcare professionals (HCPs) to administer injectable medication in the intended use environments without preventable use errors. METHODS: Patients with gMG, caregivers, and HCPs were randomized to the PFS-SD or AI. No training on either device use was provided. The safety and usability of both devices were assessed using simulated use scenarios, knowledge-based questions, and participant feedback. RESULTS: The simulated use scenario was successfully completed by 94% and 96% of participants using the PFS-SD and AI, respectively, while 90% and 87% of participants, respectively, successfully administered their assigned dose. Device acceptability was reported by 93% and 100% of patient participants for the PFS-SD and AI, respectively. CONCLUSION: This HF validation study demonstrated that the PFS-SD and AI are safe and effective for use by patients with gMG, caregivers, and HCPs in the intended use environments.