Optical phantom for resolution evaluation of non-contact corneal microscopy techniques
Year: 2026
Type: article
Abstract: We present the development of a corneal phantom designed to enable resolution assessment for non-contact corneal microscopy. The phantom was fabricated by casting a UV-curable epoxy mixed with 400-nm-diameter gold nanoparticles into a polydimethylsiloxane (PDMS) mold with a curved surface to achieve a cornea-like geometry. The radius of curvature of the final manufactured phantom was 8.14–8.17 mm with a refractive index of 1.39, similar to the adult human cornea. The optical phantom was used to evaluate the resolution of a prototype non-contact confocal microscope, the portable in vivo confocal ophthalmoscope (PICO). PICO images of the optical phantom allowed for evaluation of the resolution change as a function of depth.
Source: Biomedical Optics Express
DOI: 10.1364/boe.604912
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: gold
Estimated APC paid: $1,913
OpenAlex ID: W7167627371