Effect of a flattening filter on the surface and build-up doses for high-energy photons produced with a varian TrueBeam STx linear accelerator
Year: 2026
Type: article
Abstract: Background: The aim of thise study was to measure and determine the surface and build-up region doses at different field sizes on the central axis of 6 and 10 MV with flettening filter (FF) and without flettenin filter (FFF) high-energy photon beams, which are commonly used in clinical radiotherapy applications, and to reveal the difference between FF and FFF beams.Material and Methods: The dose measurements were made using a parallel-plate ion chamber and a water-equivalent solid water phantom for a field size of 5 × 5, 10 × 10, and 20 × 20 cm 2 , with a source skin distance of 100 cm.Results: The relative surface doses for 6 MV FF high-energy photons are 10.9% ±0.007%, 16.4%±0.010%and 26.9%±0.014%for field sizes of 5x5, 10x10 and 20x20 cm 2 .For 6 MV FFF they are 14.7%±0.010%,19.7%±0.010%and 27.2±0.009%respectively.The relative surface doses for 10 MV FF high-energy photons are 6.3% ±0.010%, 12.0%±0.007%and 23.7%±0.010%for field sizes of 5x5, 10x10 and 20x20 cm 2 .For 10 MV FFF they are 9.7%±0.010%,13.8%±0.007%and 20.1%±0.007%respectively.Conclusion: It was observed that the relative surface doses increased in FFF for 6MV compared to FF photon beams, and increased in 5x5 and 10x10 cm 2 field sizes and decreased in 20x20 cm 2 field size for 10MV photon beam.These results may have important implications in radiotherapy applications where skin dose is important.
Institution: Istanbul Medipol University
Language: English
Subfield: Radiation
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: hybrid
OpenAlex ID: W7172252722