Non-invasive blood glucose detection system using combined visible and near-infrared spectral data

Year: 2026
Type: article
Abstract: This pilot study evaluated non-invasive estimation of continuous glucose monitoring (CGM)-reported glucose from fingertip visible–near-infrared (VIS–NIR) reflectance in ten adults. Optical measurements (approximately 12 per minute) were retained as separate 70-feature vectors and paired with 1-minute FreeStyle Libre 2 CGM readings checked approximately 10 min afterward using clock-based timing; multiple optical records therefore shared one CGM label. The dataset contained 6,816 optical samples. Seven regression models were compared using sample-level five-fold and subject-independent leave-one-subject-out (LOSO) cross-validation. Under LOSO, Random Forest achieved MAE 11.248 mg/dL, RMSE 17.326 mg/dL, MARD 10.379%, and R² 0.767. In Clarke Error Grid analysis against CGM values, 82.40% and 85.30% of predictions were in Zones A and A + B, respectively. LOSO DNN classification accuracies were 92.74% and 92.43% for 30 and 15 mg/dL categories. These findings support feasibility but indicate inter-subject variability. The manually timed 10-minute offset was approximate and not an individually validated physiological lag correction. No independent capillary or laboratory blood glucose reference was available; thus clinical blood glucose measurement accuracy or safety was not established.