Entrapment of a mechanical sheath during lead extraction: Staged bailout with coaxial steerable sheath-through-laser-through-an outer sheath construct and internal snare
Year: 2026
Type: article
Abstract: Transvenous lead extraction is established for chronic leads with malfunction or infection, with class I and IIa indications defined in the 2017 Heart Rhythm Society expert consensus statement.1 The TightRail rotating dilator sheath (Philips/Spectranetics, Colorado Springs, CO) employs a bidirectional crown-type rotating blade shielded within a flexible polymer shaft, with reported clinical success rates of 89–97%.2-5 A distinctive failure mode — mechanical entrapment of the cutting edge onto the lead body, in which the sheath becomes immobile in both directions and the sheath and lead can only be moved as a single complex — has been reported in the FDA Manufacturer and User Facility Device Experience (MAUDE) database6 and as a recent abstract,7 but published procedural detail on bailout remains limited.
Source: HeartRhythm Case Reports
Author: Mohamed Shokr
Institution: Eastern Maine Medical Center
Language: English
Subfield: Cardiology and Cardiovascular Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: gold
Estimated APC paid: $1,100