Work

Prospective motion correction in structural imaging using servo navigation: pushing for high update rates and high resolution

Year: 2026
Type: conference-paper
Abstract: Motivation: Head motion can degrade image quality and bias quantitative metrics in structural imaging.Goal: To reduce motion artifacts in high-resolution structural imaging (MPRAGE and 3D-TSE) using servo navigation, and to investigate whether the geometry update rate can be increased beyond the typically long TR of these sequences. Approach: Repeated servo navigators were inserted before the inversion/excitation pulse of an MPRAGE/3D-TSE sequence to prospectively correct motion every TR (before-ET). This approach was validated in instructed motion experiments and further tested in highresolution scans. In another experiment, navigators were distributed across the MPRAGE echo train to increase the geometry update rate (within-ET), and this method was likewise evaluated in instructed motion experiments.Results: Residual motion artifacts of instructed fast in-train motion were more effectively reduced by the rapid within-ET correction compared to the slower before-ET approach. In highresolution (0.4mm iso.) experiments, image quality was preserved under involuntary motion using the before-ET correction.Impact: This work presents a novel application of servo navigation in structural imaging and demonstrates its potential to correct involuntary motion in high-resolution scans as well as rapid motion occurring during the echo train.