Work

Comparison between multiphysics simulation and experimental characterization of gradient field-induced vibrations in implantable medical devices

Year: 2025
Type: conference-abstract
Abstract: The interaction between the static magnetic field (B0) and the time-varying magnetic fields gradients (dB/dt) induces vibrations in the conductive components of active implantable medical devices (AIMDs) during MRI sequences. Vibrations can affect the operation of the AIMD and the integrity of the surrounding tissues. While recent approaches have investigated these vibrations through simulation to assess worst-case orientations and impact of tissue attenuation, they suffer from a lack of experimental validation under similar mechanical boundary conditions. This study aims to compare experimentally measured gradient-induced vibrations inside an MRI scanner with results obtained from Multiphysics Finite Element simulations.
Source: HAL
Language: English
Field: Medicine
Open Access status: green
OpenAlex ID: W7205227632
Cites:
Cited by: 1
Related to: 10