Low-cost 300 mT Benchtop MRI Scanner and Custom MR Bioreactor for real-time multi-parametric assessment of 3D-bioprinted biological tissue constructs
Year: 2024
Type: conference-abstract
Abstract: Advancements in tissue engineering and 3D bioprinting have increased the need for imaging tools that provide more than just morphological information. MRI offers multi‑parametric characterization, but conventional scanners are expensive and not easily accessible. We present a low‑cost 300 mT benchtop MRI scanner coupled with a custom, autoclaveable MR bioreactor that enables real‑time, multi‑parametric assessment of 3D‑bioprinted biological tissue constructs. Initial tests show high‑resolution morphological images of a 3D‑printed scaffold, achieving 350 μm × 350 μm in‑plane resolution in 20 minutes. This platform provides morphological and multi‑parametric data at low cost, paving the way for broader use in tissue engineering laboratories and educational settings.
Source: HAL
Language: English
Subfield: Radiology, Nuclear Medicine and Imaging
Field: Medicine
Domain: Health Sciences
Open Access status: green
OpenAlex ID: W7207081744