Accuracy in Determination of Rice Constituent Contents Using Near-infrared Transmission Spectroscopy and Improvement in the Accuracy
Year: 2002
Type: article
Abstract: The accuracy in determination of rice constituent contents using near-infrared transmission (NIRT) spectroscopy was validated. A commercial NIRT spectrometer, which was considered to be applicable to practical use for quality inspection at a rice grain elevator, was used in this study. The accuracy of the spectrometer improved after modifying light filters and calibration models. In the determination of moisture content, the coefficient of determination (r2) was 0.98, standard error of prediction (SEP) was 0.13%, and bias was -0.02%. In the determination of protein content, r2 was 0.90, SEP was 0.17%, and bias was 0.01%. These results proved that the spectrometer was sufficiently accurate to be used instead of reference analysis for measuring moisture content and protein content of brown rice. The spectrometer could be also used with sufficiently high accuracy at a grain elevator for samples in a temperature range of 10°C to 30°C and for samples with a rough rice percentage of less 10%.
Source: 農業食料工学会誌
Institution: Hokkaido University
Language: English
Subfield: Analytical Chemistry
Field: Chemistry
Domain: Physical Sciences
SDG: Zero hunger
Open Access status: green
OpenAlex ID: W2517512746