Spect定量化における面線源サイズの変化による校正係数の変動についての検討
Year: 2025
Type: article
Abstract: In recent years, quantitative evaluation in Single photon emission computed tomography/ Computed tomography (SPECT/CT) has become possible, and the system planar sensitivity method is one of the methods for calculating the calibration coefficients required for quantification. In this study, we examined the variation of the calibration coefficient in the system plane sensitivity method by changing the disk source size and the effect on the quantitative accuracy. We prepared five types (3.0 cm, 6.5 cm, 11.5 cm, 14.5 cm, and 25.3 cm) of disk source sizes and evaluated the variation of the calibration coefficients by comparing the system planar sensitivity obtained by planar collection. In addition, the quantitative accuracy was evaluated by setting Volume of interest (VOI) at 8 locations on the NEMA IEC body phantom, converting the count to radioactivity concentration, and calculating the relative error. The system planar sensitivity tended to decrease as the size of the disk source increased, and the value was highest when the disk source size was 3.0 cm. The relative error of quantitative accuracy was negative at 3.0 cm, and there was little variation at 11.5 cm or more. It was suggested that stable quantitative values could be obtained if the disk source size was between 11.5 cm and 25.3 cm.
Source: 新潟大学保健学雑誌
Language: Japanese
Subfield: Radiology, Nuclear Medicine and Imaging
Field: Medicine
Domain: Health Sciences
Open Access status: green
OpenAlex ID: W7208548417