Successful Percutaneous Retrieval of an Embolized MitraClip From the Right Subclavian Artery
Year: 2026
Type: article
Study design: Case Report
Abstract: Single leaflet device attachment (SLDA) after mitral transcatheter valve edge-to-edge repair (M-TEER) has been reported in up to 5% of patients. However, complete embolization of the M-TEER device is exceedingly rare. Device retrievals in all previously reported cases were completed by vascular surgery cut down at the end. In this case, we are presenting a complete percutaneous retrieval of M-TEER device from the right subclavian artery, which indicates the importance of procedural planning and appropriate equipment selection.
DOI: 10.1002/ccd.70860
Institutions: University of Baltimore, University of Maryland, Baltimore