Abstract: Frequency Mixing Magnetic Detection (FMMD) enables magnetic biosensing by probing the nonlinear dynamic response of magnetic nanoparticles (MNPs) under dual-frequency (f1, f2) excitation.Known Brownian-relaxation dominated MNPs BNF-Dextran 80 nm were studied in media with varying viscosities.The complex signal components at the mixing harmonics ๐‘“ 1 + 2 โ€ข ๐‘“ 2 and ๐‘“ 1 + 4 โ€ข ๐‘“ 2 were analyzed as a function of the low frequency excitation field, and the harmonic ratio ๐‘… = ๐ด (๐‘“ 1 +2๐‘“ 2 ) /๐ด (๐‘“ 1 +4๐‘“ 2 ) was evaluated as a sensitive metric.The measured ratios show a pronounced dependency on viscosity and field amplitude.The complex-plane trajectories additionally reveal viscosity-dependent changes in both amplitude and phase.These findings show that harmonic-ratio analysis in FMMD can provide environment-sensitive readout beyond conventional single-harmonic amplitude detection.The approach is therefore promising for viscosity-sensitive biosensing, where target binding or matrix changes modify the local hydrodynamic environment of the MNPs.