Kardiotokografik signallarga signal sifatiga moslashuvchan dastlabki ishlov berish algoritmi va dasturiy moduli
Year: 2026
Type: article
Abstract: Missing values and artifacts in cardiotocography signals directly affect the outcome of preprocessing, while fixed thresholds do not account for the local characteristics of the signal. This paper proposes an algorithm that combines a signal quality vector, adaptive artifact detection based on the local median and MAD, and the selection of a reconstruction strategy according to gap length and context quality. The architecture of its software module was designed, and a Python prototype was implemented. In a synthetic protocol based on reference segments from the CTU-UHB database, the F-measure of artifact detection at 10% signal loss increased from 0.75 to 0.96 on 520 test instances drawn from 157 test recordings. The reconstruction error on the same gap samples decreased from 1.94 to 1.77 bpm, and the error over all corrupted samples decreased from 4.55 to 2.19 bpm. The computation time per sample was several orders of magnitude shorter than the sampling interval.
Language: English
Topic: ECG Monitoring and Analysis
Subfield: Cardiology and Cardiovascular Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: green
OpenAlex ID: W7220916321