Pacing during TAVR through a coronary wire in the coronary artery and left ventricular wire
Year: 2026
Type: article
Study design: Case Report
Abstract: The conventional pacing-over-the-wire technique in transcatheter aortic valve replacement (TAVR) results in high impedance and voltage thresholds, which may cause loss of capture and valve embolisation. Our team reported that coronary wire pacing, in which a coronary wire is placed in the left coronary artery (anode) and a left ventricular wire (cathode), is associated with extremely low impedance and pacing voltage thresholds in patients who are at high risk of coronary obstruction. Moreover, coronary wire pacing is simple and reliable. In this study, we illustrate the coronary wire pacing technique in a patient to provide a new approach to wire pacing in TAVR.
Authors: Ruisong Ma, Lili Zhang, Wang Liao, Sheng Wang
Institutions: Hainan General Hospital, Hainan Medical University
Language: English
Subfield: Cardiology and Cardiovascular Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: gold
Estimated APC paid: $3,801