Mechanical and Hemodynamic Effects of Balloon-Expandable and Self-Expanding Stents in a Saccular Abdominal Aortic Aneurysm Model: Fluid–Structure Interaction Analyses

Year: 2026
Type: article
Abstract: Objectives: The aim of this study was to compare the effects of balloon-expandable and self-expanding stents on vessel wall mechanics and hemodynamics in a patient-specific saccular abdominal aortic aneurysm model. Methods: A patient-specific aneurysm model was constructed using computed tomography images, and balloon-expandable and self-expanding stent models were created. Aortic wall stress was evaluated by contact analysis. Fluid-structure interaction analyses were performed at 5 representative phases selected from the cardiac cycle. Mean intra-aneurysmal flow velocity, wall shear stress (WSS), pressure loss, and energy loss were compared. Results: The balloon-expandable stent produced greater localized aortic wall stress than the self-expanding stent, whereas wall stress was more uniformly distributed with the self-expanding stent. Both stents reduced the mean intra-aneurysmal flow velocity, WSS, pressure loss, and energy loss compared with the no-stent model. However, greater reductions in WSS, pressure loss, and energy loss were noted with the balloon-expandable stent than with the self-expanding stent. Conclusions: Balloon-expandable and self-expanding stents exhibited different effects on vessel wall mechanics and hemodynamics. The findings indicate the importance of evaluating both hemodynamic and mechanical effects when assessing stent-vessel interactions.
Language: English
Subfield: Surgery
Field: Medicine
Open Access status: gold
Estimated APC paid: $226
OpenAlex ID: W7223249617
FWCI:
Cites:
Cited by:
Related to: 10