Autonomous paper valves using PDMS vacuum pumps as actuators for colorimetric chrono-sampling
Year: 2026
Type: article
Abstract: a thin polymeric membrane. The pump-membrane system was experimentally characterized and used to parameterize numerical simulations, allowing predictive optimization of device performance. As a proof of concept, we demonstrate autonomous monitoring of culture medium pH at four successive time points separated by 45 minutes, with the temporal pH profile obtained from a single image acquired at the end of the experiment. This approach can be readily extended to other sampling intervals and sensor types, leveraging the inherent tunability of PDMS-based systems. Overall, this autonomous chrono-sampling platform enables low-cost, instrument-free, and temporally controlled assays, with potential applications in point-of-care diagnostics, cell culture, environmental monitoring, and time-resolved biochemical analyses.
Source: Lab on a Chip
DOI: 10.1039/d6lc00389c
Institutions: University of the Basque Country, Ikerbasque
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: hybrid
OpenAlex ID: W7225079111