Identifying sequencing artifacts in low-diversity, short-insert aptamer libraries on the Illumina MiSeq platform

Year: 2026
Type: article
Abstract: ∼1% in other samples) which resulted in overestimation of sample concentration and, in turn, decreased sequencing yield. Sequencing runs displayed highly variable pass rates (6-93%), likely due to short sequences outcompeting longer ones and assignment errors of neighboring clusters with identical early-cycle bases. Unexpected duplicates in template samples suggest that adapter-ligated daughter strands and optical misidentifications contribute to artificial duplication. These duplication artifacts are of particular importance in amplicon sequencing for which copy-number analysis is crucial. These findings highlight the need for refined ligation conditions, improved purification strategies, and enhanced bioinformatic filtering to ensure sequencing accuracy in aptamer screening library workflows on the Illumina platform.