SERS–LFIA technology based on Au@Ag nanospheres for simultaneous detection of carbapenemases KPC and NDM
Year: 2026
Type: article
Abstract: . The coefficients of variation for repeatability testing at low, medium, and high antigen concentrations were all less than 15%. The cut-off values for KPC and NDM were determined to be 846 and 784, respectively. Using PCR results as the gold standard, the diagnostic sensitivity and specificity of this method were both 100%, demonstrating a 100% agreement rate with commercial colloidal gold detection kits. The SERS-LFIA detection system may provide technical support for rapid identification of carbapenemase types.
Source: RSC Advances
DOI: 10.1039/d6ra08505a
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: gold
Estimated APC paid: $1,042
Award: G1653046035