Design and comparison of custom cranial prostheses made of titanium and polyether ether ketone using the finite element method
Year: 2026
Type: article
Abstract: Cranioplasty is a surgical procedure for reconstructing cranial defects. This study compared the biomechanical performance of patient-specific cranial implants made of Ti6Al4V and PEEK using finite element analysis. A 3D model was reconstructed from CT data of a 19-year-old male with a right parietal defect. Static loads (50–400 N) were applied at three anatomical regions. PEEK implants exhibited higher von Mises stresses than titanium implants, while transferring more load to the surrounding bone. Stress concentration occurred in Region 3 under loads >200 N. Titanium provided greater protection to bone at the cost of stress shielding, whereas PEEK offered more physiological load sharing. Results highlight the importance of material selection based on defect location and expected loading conditions.
Authors: Cyrus Ahmadi Toussi, Reza Ghasemi
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: gold
Estimated APC paid: $2,590
OpenAlex ID: W7228438063