High dose-rate response and individual correction of commercial electronic personal dosimeters for nuclear emergency radiation protection
Year: 2026
Type: article
Abstract: Reliable active dosimetry is essential during nuclear and radiological emergencies, yet the response of deployed electronic personal dosimeters (EPDs) is rarely verified for every unit in the 1– 7 S v h − 1 regime. Type testing establishes model-level compliance, whereas emergency use depends on individual instruments. We characterised ten commercial silicon-diode EPD units of the same IEC 61526-certified model in a traceable 137 Cs reference field. The units showed high repeatability ( 𝐶 𝑉 < 2 % ) but unit-dependent dose-rate bias, from −9.3% to +7.7% at 7 S v h − 1 ; one unit over-responded progressively, which a single dead-time correction cannot reproduce. A device-specific second-order polynomial correction reduced the mean RMSE from 0.187 to 0 . 0 1 1 S v h − 1 and held the residual instrumental non-linearity within ± 1 . 5 % . Applying the first-session coefficients to an independent second session, and predicting intermediate dose rates not used in fitting, both kept residuals within ± 1 . 5 % , confirming temporal reproducibility and out-of-sample validity. Bare-device directional measurements showed up to ≈ 1 9 % lateral response reduction, supporting an optional conservative planning margin ( 𝑆 𝐹 ≈ 1 . 2 ). All corrections were established for a single 137 Cs spectrum and must not be extrapolated to other photon energies without re-calibration. Individual pre-deployment characterisation is a practical complement to model-level type testing for EPDs in nuclear emergency radiation protection.
Institutions: Inje University, Korea Atomic Energy Research Institute
Language: English
Subfield: Radiation
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: gold
Estimated APC paid: $200
OpenAlex ID: W7234127203