Handheld digital biopsy device for intraoperative assessment of surgical margins during partial nephrectomy

Year: 2026
Type: preprint
Abstract: Partial nephrectomy requires reliable discrimination between renal carcinoma and adjacent renal parenchyma to achieve negative surgical margins while preserving functional tissue. In this study, we developed a handheld digital biopsy (HDB) device integrating impedance and force sensors for intraoperative surgical margin assessment. The impedance sensor incorporating bipolar electrodes was fabricated using a microelectromechanical systems (MEMS) process. Its discrimination capability was evaluated using solutions with different conductivities. To generalize the sensor output, a calibration curve (R² = 0.99) was established to estimate conductivity from impedance measurements at the optimal frequency, at which the largest impedance difference was observed. Subsequently, electrical impedance was measured in ex-vivo human papillary renal cell carcinoma (pRCC), clear cell renal cell carcinoma (ccRCC), and their corresponding adjacent renal parenchyma (pAP and cAP) under a constant contact force. The impedance of both carcinomas was significantly lower than that of their corresponding parenchyma across the measured frequency range (p < 0.05). The estimated conductivities of pRCC, pAP, ccRCC, and cAP were 0.16 ± 0.03, 0.01 ± 0.002, 1.03 ± 0.05, and 0.79 ± 0.09 S/m, respectively (p < 0.0001). These results demonstrate the feasibility of the HDB device for intraoperative renal carcinoma discrimination.