Modeling and Experimental Verification of the Temperature Dependence of Short-Period Seismometer Response, Using Pole-Zero Fitting
Year: 2026
Type: article
Abstract: Electrical calibration has long been a technique to verify the response of a seismic sensor in the lab, upon installation in the field, and with the passage of time. Subtle effects, such as temperature susceptibility, can be missed, or dismissed as random error, if the response data are not parameterized to capture the behavior of key components of the instrument. We develop a simple model that links the poles, zeros, and gain of the transfer function of a short-period seismometer to specific sensor components with inherent temperature susceptibility. We present a robust method of fitting poles and zeros to the results of a broadband calibration. We present the results of a series of calibrations of short-period S-13 seismometers in a surface vault near the Yellowknife Seismic Array, at various temperatures, and use them to validate the model. We discuss the implications for routine calibration of sensors in permanent seismograph networks. The transfer function fitting procedure presented is applicable to all seismic sensors, whereas the model of temperature dependence is applicable to any passive seismometer.
Source: Seismological Research Letters
DOI: 10.1785/0220260110
Authors: Nick Ackerley, Ziad Gias
Institution: Natural Resources Canada
Language: English
Subfield: Geophysics
Field: Earth and Planetary Sciences
Domain: Physical Sciences
Open Access status: hybrid