A Method to Reduce the Factors Affecting Iodine Enhancement for Computed Tomography Contrast Optimization

Year: 2026
Type: article
Abstract: PURPOSE: Iodine contrast enhancement is assessed using computed tomography (CT) number, though this value is not proportional to iodine concentration. This study develops a metric for iodine enhancement that is directly proportional to iodine concentration. MATERIALS AND METHODS: Using studies previously reported in the literature, we parameterized the effect to create a correction for tube potential and patient-size-specific beam hardening on iodine CT number decrease. To validate the correction, we performed a study using different iodine concentrations, phantom sizes, and tube potentials. We also applied the correction to patient pulmonary angiogram data. Patients were grouped by tube potential [80, 100, 120, 140] (kV). Quantitative iodine contrast enhancement (Q-ICE) (ie, the difference between noncontrast and contrast-enhanced CT number) of the pulmonary artery (PA) was calculated. Pearson correlation coefficients are reported to characterize trends in PA Q-ICE across weight before and after correction. RESULTS: The range of CT numbers measured for [10, 15] mg/mL iodine inserts was [290, 433] HU before correction, and [24, 36] HU after correction. Before correction, median PA Q-ICE enhancement was [662, 490, 263, 175] HU for [80, 100, 120, 140] kV, with a range of 487 HU observed across all measurements. After correction, the median PA Q-ICE enhancement was [366, 373, 285, 249] HU, with a range of 124 HU. Pearson correlation coefficients significantly (P < 0.05) decreased after correction from -0.67 to -0.37. CONCLUSIONS: This study demonstrated a method to normalize effects of tube potential and beam hardening on CT number, yielding a metric proportional to iodine concentration.
Language: English
Open Access status: hybrid
OpenAlex ID: W7234475838
FWCI:
Cites: 28
Cited by:
Related to: 10