Highly Superficial Flexible Coil Array for 1.5T MRI: Diagnostic Quality Evaluation Compared to Conventional Technology
Year: 2026
Type: article
Abstract: OBJECTIVE: To assess the imaging performance, thermal behavior, and patient acceptance of a novel 3-channel flexible coil array for 1.5T MRI (64 MHz), compared to a conventional 16-channel flexible coil. METHODS: A lightweight array with three 60 mm copper loops and a compact 32×15 mm electronics board was developed, integrating tuning, matching, decoupling, RF filtering, and preamplification. Bench tests evaluated Q-factor ratio, SNR degradation under bending, phantom-based SNR, and thermal dissipation. Clinical evaluation involved hand imaging of 10 healthy volunteers (22.9 ± 0.3 years) using Cor-T1 and Ax-T2FS sequences in two positions (superman and supine). Image quality was rated by two radiologists across five tissue types using a 5-point Likert scale. Patient comfort and anxiety were assessed via post-scan questionnaires. RESULTS: In vitro tests showed a 1.6-fold SNR improvement for superficial regions (≤10 mm depth) and approximately 40% SNR loss under extreme bending, confirming robust decoupling and conformability. Radiologists reported significantly better image quality for Ax-T2FS with the flexible array, while Cor-T1 showed no significant difference. Patient comfort was markedly higher with the flexible array (85% vs. 45%), suggesting ergonomic advantages that may reduce motion artifacts. CONCLUSION: The proposed flexible 1.5T coil array enhances superficial SNR and patient comfort without compromising diagnostic quality. SIGNIFICANCE: High-density arrays of small coils can deliver clinical benefits at 1.5T, supporting their use for improved imaging in anatomically challenging regions.
Language: English
Subfield: Radiology, Nuclear Medicine and Imaging
Field: Medicine
Domain: Health Sciences
Open Access status: hybrid
Estimated APC paid: $2,645
OpenAlex ID: W7234898198