Beyond Mean Bias in Automated ECG Interval Measurement: A Record-Level Audit in PTB-XL+

Year: 2026
Type: article
Abstract: This archive provides the submission manuscript, algorithm specification, deidentified derived analysis tables, figures, analysis manifest, and core Python scripts for a record-level audit of automated ECG interval measurements in PTB-XL+. The custom Tp-e delineator returned accepted outputs for 102 of 160 ECGs and failed all four algorithm-performance acceptance criteria. Among accepted outputs, substantial disagreement with the reader-derived human reference remained. The archive separates explicit non-output from error among retained outputs and includes calibration-overlap sensitivity analyses. Interpretation boundary: the evaluated delineator must not be used for confirmatory clinical-effect analyses. The package does not establish clinical risk thresholds or treatment effects. No credentials, private sample-mapping key, local reader-source CSVs, or unnecessary raw-waveform copies are included. PTB-XL and PTB-XL+ remain subject to their upstream terms.