Development and Performance Evaluation of an Open-System Real-Time Quantitative PCR Assay for Decentralized Hepatitis B Viral Load Testing in Indonesia
Year: 2026
Type: preprint
Abstract: Background Indonesia remains one of the countries with the highest burdens of chronic hepatitis B (CHB) in Southeast Asia. Despite this substantial disease burden, a major gap in access to quantitative hepatitis B virus (HBV) DNA testing potentially delays the diagnosis and treatment of CHB. This study aimed to develop and validate an open-system quantitative real-time PCR (qRT-PCR) assay, the INDIGEN HBV DNA qRT-PCR for detecting and quantifying HBV DNA viral load. Methods The assay was designed to target the S region of HBV and calibrated against the HBV World Health Organization (WHO) International Standard. Analytical performance was evaluated for linearity and limit of detection (LoD). A total of 122 HBsAg-positive samples from CHB patients were used to assess the clinical performance of our assay compared to the Xpert® HBV Viral Load (Xpert). Results The assay demonstrated excellent linearity across 1.0–9.0 log₁₀ IU/mL (R² = 0.9964) with a LoD of 14.8 IU/mL. Compared to Xpert, the developed assay showed 100% specificity and 99.05% sensitivity with strong correlation based on Deming regression analysis. Conclusions The INDIGEN HBV DNA qRT-PCR assay demonstrated robust evaluation performance comparable to the Xpert, supporting its implementation as a reliable open-system assay for decentralized HBV viral load testing in Indonesia.
Source: Research Square
Language: English
Topic: Hepatitis B Virus Studies
Subfield: Epidemiology
Field: Medicine
Domain: Health Sciences
Open Access status: green
OpenAlex ID: W7235181954