Work

Unsupervised surface defect detection and localization of transparent vacuum blood collection tubes using region-aware reconstruction and expert fusion

Year: 2026
Type: article
Abstract: Automated inspection of transparent vacuum blood collection tubes is complicated by specular reflections, heterogeneous endpoint structures and highly elongated regions of interest. An unsupervised region-aware reconstruction framework is developed using native-coordinate patches, three jointly trained reconstruction experts and region-prior-guided soft fusion. A Regional Multiscale Excess Score (RMES) calibrates normal regional residual tails for image-level detection while preserving the original pixel anomaly map. Evaluation on 100 abnormal ROIs and a reserved pool of 145 normal ROIs across three training seeds yielded Image AUROC of 0.9868 ± 0.0027, macro Pixel AP of 0.8358 ± 0.0710 and micro fixed-threshold Pixel F1 of 0.2499 ± 0.0052. Controlled fusion comparisons and same-map scoring experiments further assess the contributions of expert reconstruction and regional anomaly-score calibration.
Language: English
Open Access status: bronze
OpenAlex ID: W7235240153
FWCI:
Cites: 35
Cited by:
Related to: 10