Image quality and dose performance of photon-counting CT in multiple myeloma: a comparative clinical study
Year: 2026
Type: article
Study design: Observational Study
Abstract: PURPOSE: To compare image quality and radiation-dose indices between photon-counting computed tomography (PCCT) and conventional energy-integrating detector CT (EID-CT) examinations in patients with multiple myeloma. METHODS: This retrospective intraindividual study included 47 patients with non-contrast whole-body PCCT examinations (NAEOTOM Alpha, Siemens Healthineers) and prior EID-CT examinations (SOMATOM Sensation 64, Siemens Healthineers). Examinations were performed using the respective clinical protocols, which differed in acquisition and reconstruction settings. Quantitative image analysis included region-of-interest measurements in the lungs, spleen, muscle, and adipose tissue. Image noise and signal-to-noise-ratio (SNR) were calculated. For qualitative assessment, two radiologists independently evaluated randomized and anonymized PCCT and EID-CT scans of the same patients using a 5-point Likert scale to rate overall image quality, sharpness, noise, and osteolytic lesion conspicuity. Qualitative ratings were compared separately for each reader using two-sided paired Wilcoxon signed-rank tests. Differences in CT-dose-index, dose-length-product, and quantitative image metrics were analyzed using two-sided paired t-tests. RESULTS: Quantitative analysis demonstrated significantly lower image noise in PCCT compared to EID-CT in the spleen, muscle, and adipose tissue (p < 0.0001), but not in lung tissue (p = 0.27). SNR was significantly higher with PCCT in all tissues except the lungs (p < 0.0001; lungs: p = 0.22; adipose tissue: p = 0.02). Qualitative assessment showed that PCCT was rated significantly better than EID-CT for overall image quality, image sharpness, noise, and osteolysis conspicuity (p < 0.0001 for all categories). Radiation dose parameters (CTDIvol and DLP) were significantly lower with PCCT than with EID-CT (p < 0.0001). CONCLUSIONS: In non-contrast whole-body imaging of patients with multiple myeloma, the evaluated PCCT system provided superior image quality and significantly lower radiation-dose indices than the historical EID-CT system. These findings reflect the combined performance of the imaging systems and their respective acquisition and reconstruction settings.
Authors: Matthias Michael Woeltjen(Data curation, Formal analysis, Investigation, Visualization, Writing – original draft), Samir Abu Salim(Data curation, Investigation), Denise Schoenbeck, Saher Saeed(Writing – review & editing), Nina Pauline Haag(Writing – review & editing), Julius Henning Niehoff(Writing – review & editing), Jan Robert Kroeger(Writing – review & editing), Jan Borggrefe(Formal analysis, Supervision, Writing – review & editing), Alexey Surov(Conceptualization, Supervision, Writing – review & editing)
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: gold
Estimated APC paid: $4,390
Funder: Ruhr-Universität Bochum
OpenAlex ID: W7235260435