Monte Carlo assessment of a D–T neutron generator-based beam shaping assembly for BNCT: moderator effects on neutron spectrum and dose

Year: 2026
Type: article
Abstract: This study presents a Monte Carlo parametric assessment of a multilayer beam shaping assembly (BSA) for boron neutron capture therapy using a 14.1 MeV D–T neutron source. A validated reference calculation was retained for the principal beam-quality quantities, while additional controlled calculations were used to reassess material substitution, finite-source and aperture effects, and one-at-a-time layer-thickness sensitivity. Under a common computational framework, AlF 3 provided the highest epithermal response and the lowest fast-to-epithermal ratio among the tested moderator substitutions. Paraffin and light water reduced the epithermal component by approximately 74% and 73%, respectively, relative to AlF 3 , whereas D 2 O and polystyrene produced intermediate and similar responses. The validated reference configuration yielded an epithermal neutron flux of 1.1×10 9 n cm⁻ 2 s⁻ 1 and a beam directionality of 0.78; however, its fast-neutron dose per epithermal flux, 2.4×10⁻ 13 Gy·cm 2 /n, remained about 20% above the historical IAEA-TECDOC-1223 design benchmark. Additional calculations showed that the finite 1 cm-radius source did not produce a statistically resolved change relative to the point-source reference, whereas increasing the aperture from 10 to 24 cm increased the aperture-averaged epithermal response. Thickness variations revealed substantial transmission trade-offs and did not establish a unique global optimum. The 36 cm AlF 3 , 4 cm Fe, and 2.6 cm Cd dimensions are therefore reported as the reference geometry of the existing assembly. The results support continued computational refinement but do not constitute clinical qualification.
Language: English
Field: Medicine
Open Access status: gold
Estimated APC paid: $2,690
OpenAlex ID: W7235851068
FWCI:
Cites: 24
Cited by:
Related to: 10