Method Development and Validation for Morphine in Chicken Plasma by RP-HPLC for Quantitative Analysis
Year: 2026
Type: preprint
Abstract: The study focuses on the development and validation using reverse phase HPLC method for the quantitative estimation of morphine in chicken plasma, that is applicable to Forensic analysis in an accurate, precise and simple way. C18 column (250×4.6 mm, particle size 5 µm) with an isocratic method using mobile phase Methanol and 0.1% orthophosphoric acid in Milli-Q water (60:40 v/v) was used for the chromatographic separation. Morphine was extracted utilizing a liquid-liquid extraction method. The flow rate inside the column was kept at 1.0 ml/min and detected using a photodiode array detector at an appropriate wavelength of 240nm. The method was validated in accordance with ICH Q2(R2) guidelines, evaluating parameters including accuracy, precision, specificity, linearity, limit of detection (LOD) and limit of quantification (LOQ). The method showed good linearity over concentration range of 1.66–8.33 µg/ml, with a correlation coefficient (r 2 ) of 0.9991. Precision studies indicated low %CV values, confirming reproducibility. Recovery studies demonstrate acceptable extraction efficiency. The developed method is simple, rapid, cost-effective, and suitable for routine quantitative analysis of morphine in forensic.
Source: Research Square
Institution: National Forensic Sciences University
Language: English
Subfield: Toxicology
Domain: Life Sciences
Open Access status: green
OpenAlex ID: W7235882584
