Abstract: : Spherical markers are commonly used by phantombased calibration methods for X-ray CT systems. Defining the position of the marker centers is therefore crucial to estimate the geometry parameters accurately. Although marker bearing structures are often built from materials of low X-ray attenuation, they still overlap with the marker in projection images. This complicates accurate determination of the marker centers. In this work, we explore the technique of Deep Learning to extract the marker center coordinates from the calibration projections. By training a Deep Learning network for each marker center coordinate, 2D positions of the marker are derived. With simulated as well as real experiments, it is shown that the trained Deep Learning networks can be used to accurately estimate the marker positions, and hence also the geometry of the X-ray CT system.
Institutions: University of Antwerp, IMEC, University of Insubria
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Keywords: deep learning, X-ray images, phantoms
Open Access status: green