Clinical feasibility of a new fitting procedure for active bone conduction implants

Year: 2026
Type: article
Abstract: Purpose To compare, in users of Bonebridge (BB) system, the outcomes using two first fit algorithms, in an acute test setting with the patient’s individually fine-tuned fitting based on standard fitting methods. Methods 22 individuals aged 5 years and older in the BB’s indication range were included. Non-experienced patients were enrolled prospectively, and experienced BB users were included retrospectively. Results Mean sound field PTA 4 thresholds were: 41 ± 7dB for standard, 32 ± 5dB for new and 32 ± 10 dB for own fittings ( p < 0.001). Mean wor​d recognition scores at 45dB were: 22 ± 30% for standard, 71 ± 19% for new and 52 ± 40% for the patient’s own fitting condition ( p < 0.001). A ceiling effect was observed at 65dB ( p = 0.36). The average results of the speech in noise tests were: -2.29 ± 1.97 dB SNR for standard, -4.11 ± 1.53 dB SNR for new and − 3.65 ± 1.77 dB SNR for the patient’s own fitting conditions ( p = 0.002). 85.7% of test subjects preferred the new fitting formula over the standard fitting formula. Conclusions The new fitting formula demonstrated improved performance in both audiological and subjective measures in comparison to the standard. Further research is necessary to validate these findings in broader clinical settings.