Filtered high-energy X-ray generator as an alternative to cobalt-60 for total ionizing dose testing of electronics

Year: 2026
Type: article
Abstract: Total Ionizing Dose testing of electronic components and systems has traditionally relied on the use of radioactive sources. Such sources raise safety and security concerns, creating a clear need for safer alternatives in industrial, medical, military, and research applications. In this work, a methodology for irradiating electronic components using filtered high-energy X-rays is evaluated, with the objective of preserving results comparable to those obtained using existing testing standards. Simulations are performed to select an appropriate X-ray filter capable of reproducing the degradation induced by 60 Co irradiation. A tin and copper filter is identified that attenuates low-energy photons while minimizing dose enhancement effects. Several irradiation conditions are investigated. The results demonstrate that irradiation with a high-energy X-ray generator using the tin and copper filter reproduces 60 Co degradation more closely than other filter configurations. A conversion factor is applied to compare component degradation at equivalent calculated doses in SiO 2 across the different irradiation conditions.