Work

Reported Intraoperative Equipment Problems, Troubleshooting, and Operative Disposition During Robot- Assisted Nephroureterectomy: An FDA MAUDE Analysis

Year: 2026
Type: preprint
Abstract: Intraoperative equipment problems during robot-assisted nephroureterectomy (RANU) are poorly characterized. We searched FDA Manufacturer and User Facility Device Experience (MAUDE) reports received from January 1, 2015, through September 30, 2026, and included reports documenting both RANU and an intraoperative equipment problem. Related reports were linked; equipment problems and troubleshooting were analyzed per distinguishable failure, and operative disposition per operation. Of 870 unique MDRs screened, 168 were included, representing 153 distinguishable failures in 151 operations. Breakage, detachment, or visible structural damage was the most frequently coded primary manifestation (53/153, 34.6%), followed by abnormal movement, jaw operation, or jamming (30/153, 19.6%). At least one of six predefined troubleshooting actions was documented for 87 failures (56.9%), including equipment exchange in 41 (26.8%). A specific final operative path could be identified for 27 operations (17.9%). Completion was explicitly documented for 103 (68.2%), including 84 completed operations with no specified final approach; another 40 had no reported final operative disposition. These reports describe equipment problems and intraoperative responses during RANU, but cannot establish malfunction incidence or device-attributable clinical risk.
Language: English
Open Access status: green
OpenAlex ID: W7238788228
Cites:
Cited by:
Related to: