Management of failed perclose proglide suture during transfemoral TAVR: a bailout strategy

Year: 2026
Type: article
Abstract: Transcatheter aortic valve replacement (TAVR) is a minimally invasive procedure preferred for elderly patients with severe aortic stenosis. Vascular access complications, though rare, can pose significant challenges, particularly when using a single ProGlide closure device (Abbott, USA). In this case, we describe a patient in his 80s undergoing transfemoral TAVR who experienced a snapped single ProGlide suture. A systematic bailout strategy involving contralateral femoral access and balloon-assisted hemostasis was employed to achieve vascular closure successfully. This report highlights the importance of pre-procedural planning, mastery of alternative techniques, and stepwise procedural refinements to manage complications effectively. The case underscores the necessity of tailored approaches in complex vascular access scenarios to avoid surgical intervention.