Analytical agreement and classification performance of two point-of-care devices for lipid panel measurement
Year: 2026
Type: preprint
Abstract: Background Dyslipidemia is a major risk factor for cardiovascular disease. Point-of-care Testing (PoCT) devices provide rapid lipid measurements and may facilitate timely risk assessment and clinical decision-making. However, comparative evidence on their analytical agreement for classification performance, operational feasibility, and acceptability remains limited. Objective To evaluate the classification performance, analytical agreement and operational feasibility of the Mission® Cholesterol and CardioChek® Plus Analyzer PoCT devices for lipid profile assessment compared with the reference laboratory method, and to assess participant acceptability, providing evidence to support the implementation of these devices in dyslipidemia screening and monitoring programs within healthcare systems. Methods A cross-sectional study was conducted in primary healthcare settings. Paired capillary and venous blood samples were obtained to compare Mission® and CardioChek® with laboratory reference measurements of total cholesterol (TC), HDL-C, LDL-C, and triglycerides (TG). Agreement analysis and clinical classification metrics were calculated, while participant’s acceptability of finger-prick sampling was evaluated. Results Among the 279 participants with paired measurements, the mean age was 58 ± 12 years, and 79% were female. Both devices showed strong correlations with the laboratory reference method, with ICCs ranging from 0.871 to 0.920 for Mission® and from 0.898 to 0.957 for CardioChek®. CardioChek® showed greater analytical agreement for TC, HDL-C, and TG, whereas Mission® showed greater agreement for LDL-C. Classification performance varied by biomarker, with Mission® showing higher sensitivity for HDL-C and LDL-C abnormalities and CardioChek® showing better classification performance for elevated TG. Finger-prick blood collection was preferred by 51.3% of participants, compared with 16.1% who preferred venous collection, while 32.6% reported no preference; 94.3% reported no or mild finger-prick pain. Conclusion Both devices may support decentralized lipid screening in primary care, provided that biomarker-specific limitations are considered, and laboratory confirmation is recommended when clinically appropriate. Trial registration: Not applicable.
Source: Research Square
Institution: Hospital Israelita Albert Einstein
Language: English
Subfield: Medical Laboratory Technology
Field: Health Professions
Domain: Health Sciences
Open Access status: green
OpenAlex ID: W7238871293