Year: 2026
Type: preprint
Abstract: The present study explores interpolation errors at the melting point of gallium for four sub-ranges above the triple point of water covering from 0.01°C to 660.323°C in the temperature range of standard platinum resistance thermometers. A substitution of actual measurement for standard thermometers quality assessment is proposed on the grounds of error analysis. Interpolated value at the melting point of gallium by the ITS-90 interpolation from 0.01°C to 419.527°C is selected as the metric for assessing the quality of standard thermometers This approach would substantially reduce the number of measurements required at the melting point, thereby lowering measurement costs in the relevant subranges.