Development and fabrication of heat-sterilizable inhalation therapy equipment
Year: 1974
Type: article
Abstract: The development of a completely heat sterilizable intermittent positive pressure breathing (IPPB) ventilator in an effort to reduce the number of hospital acquired infections is reported. After appropriate changes in materials and design were made, six prototype units were fabricated and were successfully field tested in local hospitals. Most components of the modified ventilators are compatible with existing machines. In all but a few instances, such as installation of bacteria-retentive filters and a modified venturi, the change over from non-heat-sterilizable to sterilizable units was accomplished by replacement of heat labile materials with heat stable materials.
Author: Alexander S. Irons
Institution: Jet Propulsion Laboratory
Language: English
Topic: Nosocomial Infections in ICU
Field: Medicine
Domain: Health Sciences
Open Access status: green
OpenAlex ID: W748420939