Differential scanning calorimetry as a tool for optimizing vegetables freezing and storage conditions

Year: 2007
Type: conference-paper
Abstract: Humidity determination: The humidity of samples (%) was determined by the NP 784 1970 (data presented are averages from 4 measurements). DSC analysis: Equipment_ 823e Mettler Toledo equipped with a TSO800GC1 gas unit and a Intercooler Thermo HAAKE EK 45/ MT. Purge gas_ Nitrogen (20 ml min-1) Calibration_ Indium (m.p.156.6oC, H = 28.4J g-1) and deionised water (m.p.: 0 oC, H = 334.5 J g-1), as standard procedure. Pan_ aluminium sealed pan (ca. 40 l). An empty, sealed aluminium pan was used as the reference. Parameters definition_ Glass transition temperature: was taken as two different criteria, Tg` middle, the temperature at which the specific heat variation was half of the total value at the transition; Tg onset: the temperature of departure from the specific heat line of the glassy state before the transition; Crystallization heat (Hc): area of the peak during cooling run (1o peak); Melting heat (Hm): area of the peak during heating run (2o peak); Crystallization and melting temperature, Tc and Tm, the highest point in 1o and 2o peak, respectively. Duplicated samples were subjected to the following heating program related with the measurements needs: Differential Scanning Calorimetry as a tool for optimizing vegetables freezing and storage conditions
Authors: Gonçalves, E. M., Lopes, C., Pinheiro, J., Lourenço, J. A. A., Abreu, M., Brandão, Teresa R. S., Silva, C. L. M.
Language: English
Open Access status: green
OpenAlex ID: W756636233
FWCI:
Cites:
Cited by: 2
Related to: 10