Evaluation of a volume-ratio calibration system for vacuum gages from 10-6 to 10-3 torr

Year: 1965
Type: article
Abstract: An evaluation was made of a volume-ratio calibration system in the range from to of e r r o r ranging from about 6 percent at t o r r to 1 percent at was also made of the effects of ionization gage pumping, outgassing, ultimate pressure, and adsorption.Finally, commercial ionization gages were calibrated in the system.Large differences in calibration factors were observed between different gages along with various nonlinearity effects; however, the reproducibility of a typical ionization gage calibration curve was found to be about 3 percent.torr.An e r r o r analysis showed that the pressures could be produced with a limittorr.An analysis