Work

ELECTROCHEMICAL CHARACTERISATION OF X10CrNi 18-8 STEEL IN ARTIFICIAL PLASMA Received – Prispjelo: 2014-05-02

Year: 2015
Type: article
Abstract: Quality of applied guide wires conditions the course of treatment and its success to a great extent.In order to simulate conditions that can be found in blood vascular system samples were exposed to artificial blood plasma (T = 37 ± 1 oC) for the time of 8 h.In order to obtain information regarding physical and chemical properties of modified surface of wire made of X10CrNi 18-8 steel, electrochemical impedance spectroscopy (EIS) test and tests of chemical composition of the surface layer X-ray photoelectron spectroscopy (XPS) were made.On the ground of performed EIS and XPS tests, favourable impact of steam sterilisation process on corrosion resistance of X10CrNi 18-8 steel was observed only for the case when chemical passivation was applied prior to sterilisation.
Language: English
Open Access status: green
OpenAlex ID: W860666808
FWCI:
Cites: 8
Cited by:
Related to: 9