Work

Calculation of Full Energy Peak Efficiency of NaI (Tl) Detectors by New Analytical Approach for Cylindrical Sources

Year: 2013
Type: article
Abstract: A new analytical approach for calcu lation of the full-energy peak efficiency of NaI (Tl) is deduced. In addition, self attenuation of the source matrix, the attenuation by the source container and the detector housing materials are considered in the mathematical treat ment. Results are co mpared with those measured by two cylindrical NaI (Tl) detectors with Resolution (FWHM ) at 662 keV equal to 7.5% and 8.5%. 152 Eu aqueous radioactive sources covering the energy range fro m 121 keV to 1408 keV were used. By comparison, the calculated and the measured full-energy peak efficiency values were in a good agreement.
Language: English
Subfield: Radiation
Open Access status: diamond
OpenAlex ID: W91108085
FWCI:
Cites: 17
Cited by: 1
Related to: 10