Work

Development and validation of RP-HPLC method for estimation of eplerenone in spiked human plasma

Year: 2012
Type: article
Abstract: A rapid and simple high performance liquid chromatography (HPLC) method with a UV detection (241 nm) was developed and validated for estimation of eplerenone from spiked human plasma. The analyte and the internal standard (valdecoxib) were extracted with a mixture of dichloromethane and diethyl ether. The chromatographic separation was performed on a HiQSil C-18HS column (250 mm×4.6 mm, 5 μm) with a mobile phase consisting of acetonitrile:water (50:50, v/v) at flow rate of 1 mL/min. The calibration curve was linear in the range 100-3200 ng/mL and the heteroscedasticity was minimized by using weighted least squares regression with weighting factor 1/X.
Language: English
Field: Chemistry
Open Access status: gold
Estimated APC paid: $1,950
OpenAlex ID: W2009448641
FWCI:
Cites: 6
Cited by: 17