Development and Validation of a Reversed-Phase HPLC Method for In Vitro Loading and Release Analysis of Paclitaxel Coated Stent

Year: 2008
Type: article
Abstract: A simple, rapid and sensitive high-performance liquid chromatographic method is described for determination of Paclitaxel (PCL) from coated stent surface using a reverse phase X-terra C 18 (5 im) column at 227 nm and using acetonitrile, methanol and water in the ratio of 60:5:35 (v/v) as a mobile phase following single-step extraction from Phosphate buffer saline (PBS) pH 7.4 with dichloromethane. The assay was validated against the classical criteria and was applied to detect paclitaxel content as well as release amount from the stent surface. Sink condition of paclitaxel coated stent was maintained in PBS throughout the experiments by different criteria. It was maintained by modifying the release medium such as methanolic PBS, PBS containing DENA (N, N- Diethylnicotinamide), and 0.1% sodium azide in PBS. With the “sampling” technique, drug-loaded stent was introduced into a vessel, and release was monitored over a period of time by analyzing aliquots of release medium. In the use of agitation by means of orbital shaking incubator, the sampling seems to be easier than the continuous flow method. In the “continuous flow” technique, media is continuously circulated through a cell containing drug-loaded stent followed by analysis. The extraction method achieves a chemical separation of drug from the release media by use of a separating funnel. With all these methods, the setup and sampling techniques seem to influence in vitro release. © Society for Biomaterials and Artificial Organs (India), 20071205-20.
Language: English
Field: Chemistry
Open Access status: diamond
OpenAlex ID: W999632453
FWCI:
Cites: 5
Cited by: 2
Related to: 9